Capture
1-31
SR785 Dynamic Signal Analyzer
Use the knob to select (Playback) and press
[Enter].
Choose Playback from Capture. Since there is
captured data for both inputs, both displays start
measuring from the capture.
Press [Auto Scale A] and [Auto Scale B].
Scale the displays to show the measurements.
Press
[Start/Reset]
Start playback from the beginning of the buffer.
Since the default Playback Speed is Normal, the
entire playback takes as long as the equivalent real
time measurement. In this case, the capture
represents 7.73 seconds of data so the playback
takes 7.73 seconds as well.
At the current span (51.2 kHz), there are almost
1000 time records of captured data. Not all time
records are displayed during Normal playback
though all time records contribute to averaged
measurements. The display is updated 8 times a
second for 7.73 seconds for a total of 62 updates
during this playback.
Normal playback is limited to the real time
limitations of the equivalent real time analog input
measurement.
Press <Playback Config>
Enter the Playback Configuration submenu.
Press <Playback Speed>
Change the Playback Speed.
Use the knob to select (Every Time Rec) and
press [Enter].
Choose Every Time Record. Since the data is
stored in memory, we can choose to display every
stored time record.
Press
[Start/Reset]
Start the playback at the beginning of the buffer.
In this case, all 1000 time records are measured
AND displayed. The display still updates at 8 Hz so
playback takes about 125 seconds to complete.
If the time records are overlapped, there may be
more than 1000 measurements to display and
playback will take even longer.
Every Time Record playback is not limited by real
time considerations.
Press [Trigger]
Select the trigger menu.
Press <Trigger Source>
We'll use manual trigger to show how you can step
through each time record in the capture buffer.