Waterfall
Display
1-37
SR785 Dynamic Signal Analyzer
9. Press [Waterfall]
Select the Waterfall menu.
Press
<Display>
Change the Display.
Use the knob to select (Waterfall) and press
[Enter].
Choose Waterfall display. This shows
measurements scrolling down. Without waterfall
storage, this is purely graphical, no data can be read
from measurements other than the most recent.
Press
<Storage>
Select waterfall Storage.
Press (Save Options) and press [ENTER]
Use the knob to select (Active Meas Only)
and press [ENTER]
Use the knob to select (One Shot) and press
[Enter].
Choose One Shot to fill the waterfall buffer once
and stop.
Press <Total Count>
Change the number of measurements to store in the
waterfall buffer.
Press [5] [0] [Enter]
The Total Count is linked to both displays by
default. Entering 50 changes the total count for
both displays to 50.
10. Press [Trigger]
Select the Trigger menu again.
Press
[Start/Reset]
[Start/Reset] starts the measurement. Since the
measurement is triggered (not Free Run), nothing
happens until the first trigger is received.