4-154 Limit Testing Analysis Menu
SR785 Dynamic Signal Analyzer
Command: LSEG (?) d, i {, j, x0, y0, x1, y1}
Segment #
Select the Current Segment for editing [0..n]. The limit Type and Start (X0, Y0) and End
(X1, Y1) points for this segment are edited in this menu. The Current Segment is
identified by two triangular arrows located at the segment endpoints. These arrows are
above the Upper or below the Lower Limit Type markers.
Use <Delete Segment> to remove this segment. The numbering of the remaining
segments may change as a result. Use the Current Segment markers to identify the
correct segment before editing.
X0
Enter or adjust the X coordinate for the Current Segment Start point. The X coordinates
are assumed to be in the display units (usually Hz).
Use Zoom and Pan to expand the display when drawing small segments.
Command: LSEG (?) d, i {, j, x0, y0, x1, y1}
Y0
Enter or adjust the Y coordinate for the Current Segment Start point. The Y coordinates
are assumed to be in the display units.
Use Zoom and Pan to expand the display when drawing small segments.
Command: LSEG (?) d, i {, j, x0, y0, x1, y1}
X1
Enter or adjust the X coordinate for the Current Segment End point. The X coordinates
are assumed to be in the display units (usually Hz).
Use Zoom and Pan to expand the display when drawing small segments.
Command: LSEG (?) d, i {, j, x0, y0, x1, y1}
Y1
Enter or adjust the Y coordinate for the Current Segment End point. The Y coordinates
are assumed to be in the display units.
Use Zoom and Pan to expand the display when drawing small segments.
Command: LSEG (?) d, i {, j, x0, y0, x1, y1}