Octave
Analysis
1-23
SR785 Dynamic Signal Analyzer
Octave Analysis
This example investigates the test filter (enclosed with this manual) using Octave measurements. You
will use the SR785 source to provide a broad band source and both displays to measure the output of the
device under test.
Refer to ‘Octave Analysis’ in Chapter 2 for more about Octave Analysis measurements.
1. Press [System]
Display the System menu.
Press
<Preset>
Preset returns the unit to its default settings.
Press [Enter] to confirm Preset.
Preset requires confirmation to prevent accidental
reset. Wait until the self tests are completed.
2. Use a BNC Tee to connect the Source Output to
the filter input and the Ch1 A Input.
Connect the filter output to the Ch2 A Input.
Ch1 measures the source (filter input) and Ch2
measures the filter output.
3. Press [Display Setup]
Select the Display Setup menu.
Press <Measure Group>
Change the Measurement Group.
Select (Octave) with the knob and press [Enter].
Choose the Octave group. Both displays are now
making Octave Analysis measurements.
Press [Source], <On>.
Turn the source on.
Press [Link] [Auto Scale A]
Auto Scale DisplayA and change the scale of
DisplayB to match DisplayA.
The Octave measurement displays the output of
logarithmically spaced bandpass filters. This is not
an FFT based measurement. The last bin at the right
is a Sound Level measurement and may be
calculated independently from the octave bands.
Note that even though the source is a single
frequency sine wave, the octave display shows a
very broad peak. This is because the individual
bandpass filters are very broad, 1/3 of an octave in
this case.