4-112 Average Menus
SR785 Dynamic Signal Analyzer
measurement is not settled until the discontinuity in the input data has propagated
through the digital filters and a complete new time record has been acquired.
If the time record increment is 100%, unsettled measurements are not displayed. After a
change is made which unsettles the measurement, new data is not displayed until the
filters are settled and a complete time record has been acquired.
If the measurement is running with a time record increment less than 100% and the
measurement is unsettled, unsettled measurements may be displayed. New data is
displayed after the filters are settled and a portion of the new time record has been
acquired. For example, if the time record increment is 25%, 3 measurements are made
before a complete new time record has been acquired. These first 3 measurements have
time records which contain data from before AND after the measurement was unsettled.
These unsettled measurements are displayed in half intensity indicating that the
measurement is not settled. Once the 4th measurement is complete, the display returns to
full intensity since the measurements have completely settled time records.
If averaging is on, changes which unsettle the measurement will restart the average.
Unsettled measurements are not included in the new average. Averaging does not start
until the measurement is settled. When the time record increment is less than 100%,
‘Settle’ is displayed (instead of the number of averages) below the graph to indicate that
the unsettled measurements are not being averaged.
See also Real Time Bandwidth and Overlap.
If the Analyzer Configuration is 'Independent Channels,' each display can have its own
Time Record Increment. This entry field can be linked to both displays by using the
[Link] key.
Command: FOVL (?) d {, x}
More
This submenu selects additional averaging parameters.
Overload Reject
Select Overload Reject for both displays [Off, On].
When Overload Reject is On, time records which contain points which were overloaded
at the input are not included in the measurement average. Whenever a time record is
rejected, the ‘Reject’ indicator below the graph (next to the average type and number)
turns on briefly. This has no effect if Averaging is Off.