4-84 Swept Sine Source Menu
SR785 Dynamic Signal Analyzer
Source Ramping is Off, the source shuts off instantly in these situations. If Source
Ramping is On, the source will ramp off. When the sweep is started or re-started, the
source will ramp on for the first measurement.
The amplitude may be changed at any time during a sweep.
Command: SSAM (?) {x}
Ideal Reference
Set the Ideal Reference [0 mV .. 5000 mV]. This parameter is adjustable only if Auto
Level Reference is set to Channel 1 or Channel 2.
The Ideal Reference is the signal level that the source maintains at the Reference
Channel to within the Reference Limits. This is determined by the limitations of the
device under test or may simply be the desired signal level for the test. If the Ideal
Reference requires a source amplitude greater than the Maximum Source Level, then the
Reference Channel signal will fall below the Ideal Reference. In cases where the
required source amplitude is very small, the resolution of the source amplitude may
result in a Reference Channel signal which is not within tolerance of the Ideal Reference.
The Ideal Reference may be changed at any time during a sweep.
Command: SSRF (?) {x}
Source Ramping
Select the Source Ramping Mode [Off, On].
If Source Ramping is Off, source level changes are made instantly.
If Source Ramping is On, source level changes are made at the Source Ramp Rate.
Settling starts after the source amplitude reaches the desired level. This prevents fast
level changes from disturbing the device under test.
The swept sine source turns off whenever there are no measurements being made. This is
before the sweep is started, at the end of a single sweep or while a sweep is paused. If
Source Ramping is Off, the source shuts off instantly in these situations. If Source
Ramping is On, the source will ramp off. When the sweep is started or re-started, the
source will ramp on for the first measurement.
Command: SRMP (?) {i}
Source Ramp Rate
Set the Source Ramp Rate [0.001 V/s .. 500 V/s].
The Source Ramp Rate is the rate at which the source amplitude changes when Source
Ramping is On. If Source Ramping is Off, source amplitude changes are made instantly.
Содержание SR785
Страница 4: ...ii ...
Страница 10: ...viii ...
Страница 80: ...1 64 Exceedance Statistics ...
Страница 158: ...2 78 Curve Fitting and Synthesis SR785 Dynamic Signal Analyzer ...
Страница 536: ...5 136 Example Program SR785 Dynamic Signal Analyzer ...