2-48 Swept Sine Measurements
SR785 Dynamic Signal Analyzer
FFT Measurement
Swept Sine Measurement
Figure Chapter 2 -2 Frequency responses
The range of the FFT measurement is limited to about 80 dB. This is because the
response signal to Channel 2 contains frequency components within the filter pass band.
These components add to create amplitude peaks near 0 dBV (for a 1V chirp amplitude).
This requires the input range of Channel 2 to be set near 0 dBV even though each pass
band component is only at -30 dBV. The noise floor at full span is about -100 dBV.
Thus, the dynamic range of this measurement is roughly 70 dB (pass band to noise
floor). Vector averaging in this case improves the dynamic range to about 80 dB.
The swept sine sweep measures each frequency alone, thus optimizing each frequency
point individually. This measurement was taken using Auto Range to adjust the input
range at each frequency for the best signal to noise. When the frequency is in the stop
band, the return signal to Channel 2 is -80 dBV (or less). The input range of Channel 2 is
adjusted to -50 dBV for these frequencies eliminating the noise floor limitation. The
depth of the zero in the response as well as the shape of the stop band are clearly
resolved.