5-102 Front Panel Commands
SR785 Dynamic Signal Analyzer
Front Panel Commands
ACTD (?) i
The ACTD command sets the Active Display to display i. The parameter i
selects DisplayA (0) or DisplayB (1).
Use the ACTD command to select the active display in the Single format. In the
Single format, the inactive display is still accessible via commands.
STRT
The STRT command starts the measurement. Any average in progress is reset
and started over. If the measurement is paused, STRT starts the measurement
over.
This function is the same as pressing the [Start/Reset] key.
PAUS
If the measurement is already in progress, PAUS pauses the measurement.
If the measurement is paused, PAUS has no effect.
This function is similar to pressing the [Pause/Cont] key.
CONT
If the measurement is paused, CONT continues the measurement.
If the measurement is running, CONT has no effect.
This function is similar to pressing the [Pause/Cont] key.
UNST d
The UNST command unsettles the measurement of display d. The measurement
is not actually perturbed by the UNST command. The settling status is set to
unsettled and the full settling time of the measurement is required before the
status returns to settled.
The measurement is unsettled by changing any one of several measurement
parameters. For example, changing the input range or FFT span will unsettle the
measurement. If the signal comes from an external source and is changed in such
a way as to require the measurement to settle, it is convenient to use the UNST
command and wait for settling to finish.
This command is valid only when the Measurement Group is FFT or Octave.
SVTR d, i
The SVTR command saves display d to Trace i.