SR785 Dynamic Signal Analyzer
[Marker Center]
[Marker Center] sets the FFT Center frequency to the Marker frequency in the active
display. If the new Center frequency and Span conflict with the maximum measurement
range, 0 to 102.4(100.0) kHz, then the Span is decreased to the largest span which allows
the Marker frequency to be the center.
[Link] [Marker Center] adjusts the span of the active display and sets the Span of the
inactive display to match.
Command: MKCN d
[Marker Max]
[Marker Max] moves the Marker to the location of the maximum data value within the
active display. [Marker Max] only searches the data which is in the display. If the
maximum value occurs at more than one location, then the one closest to the left edge is
found.
If the Marker Mode (in the [Marker] menu) is Normal, [Marker Max] centers the marker
region around the maximum. The Marker Seeks Mode chooses whether the Marker seeks
the Maximum, Minimum or Mean of the data within the Marker Region.
For Harmonic or Sideband markers, [Marker Max] moves the Fundamental Marker to the
location of the maximum. [Marker Max] has no affect on Band markers.
[Link] [Marker Max] moves the Marker to the location of the maximum within both
displays.
Command: MKMX d
[Marker Min]
[Marker Min] moves the Marker to the location of the minimum data value within the
active display. [Marker Min] only searches the data which is in the display. If the
minimum value occurs at more than one location, then the one closest to the left edge is
found.
If the Marker Mode (in the [Marker] menu) is Normal, [Marker Min] centers the Marker
Region around the minimum. The Marker Seeks Mode chooses whether the marker seeks
the Maximum, Minimum or Mean of the data within the Marker Region.
For Harmonic or Sideband markers, [Marker Max] moves the Fundamental Marker to the
location of the minimum. [Marker Max] has no affect on Band markers.
[Link] [Marker Min] moves the marker to the location of the minimum within both
displays.
Command: MKMN d
[Show Setup]
[Show Setup] enters the Help system and displays the measurement setup and system
settings. Press [0] to exit Help and return to the measurement displays.