1-6 Getting
Started
SR785 Dynamic Signal Analyzer
Check that the Trigger Mode is set to Auto Arm. If the Trigger Mode is Manual Arm,
then the analyzer will only trigger once and then wait for the next Manual Arm
command.
Scaling and Ranging
Check that the inputs are not completely overloaded by using [Auto Range Ch1] and
[Auto Range Ch2].
Scale the display to show the entire range of the data with [Auto Scale A] and
[Auto Scale B].
Local
Make sure that the analyzer is not in the REMOTE state where the computer interfaces
have setup the instrument and locked out the front panel. Press the [Local/Help] key to
restore local control.
Reset
If the analyzer still seems to function improperly, turn the power off and turn it back on
while holding down the [<-] (backspace) key. This will reset the analyzer into the default
configuration. The analyzer should power on running and taking measurements.