Linking (Advanced Operation) 1-15
SR785 Dynamic Signal Analyzer
Linking (Advanced Operation)
This example investigates the test filter (enclosed with this manual) using FFT measurements. You will
use the SR785 source to provide a broad band source and both displays to measure the output of the
device under test. Display parameter linking and function linking will be explored in greater detail.
1. Press [System]
Display the System menu.
Press
<Preset>
Preset returns the unit to its default settings.
Press [Enter] to confirm Preset.
Preset requires confirmation to prevent accidental
reset. Wait until the self tests are completed.
2. Connect the Source Output to the filter input.
Connect the filter output to the Ch2 A Input.
In this example, only the filter output on Channel 2
is required.
3. Press [Source]
Select the Source menu.
Press
<Noise>
Choose Noise output. The output is random noise.
Press <On>
Turn the source on.
Press
<Type>
Adjust the Noise Type.
Select (White) with the knob and press [Enter].
This source is White Noise which extends over the
entire 0-102.4 kHz frequency range.
4. Press [Auto Range Ch2]
Let the analyzer automatically set the Input Range
to agree with the signal. Note that the Input Range
readouts at the top of the screen are displayed in
inverse when Auto Range is on.
5. Press [Display Setup]
Select the Display Setup menu.
Press
<Measurement>
Change the Measurement of the active display (A).
Select (FFT ch2) with the knob and press
[Enter].
Choose FFT spectrum of Ch2 for the measurement
in DisplayA (top). Both displays are independently
measuring the filter output spectrum.
6. Press [Input]
Select the Input Menu
Press <Analyzer Config>
Change the way the input channels are used.