Display Options Commands 5-43
SR785 Dynamic Signal Analyzer
Display Options Commands
DISP (?) d {, i}
The DISP command sets (queries) the Display Update Mode of display d. The
parameter i selects Off-Line (0) or Live (1).
Many parameters (Frequency, Measurement, etc.) are not adjustable for an Off-
Line display. Commands which set these parameters for an Off-Line display are
not valid.
DFMT (?) {i}
The DFMT command sets (queries) the Display Format. The parameter i selects
Single (0), Dual (1), or Overlay (2).
Use the ACTD command to select the active display in the Single format. In the
Single format, the inactive display is still accessible via commands.
ACTD (?) i
The ACTD command sets the Active Display to display i. The parameter i
selects DisplayA (0) or DisplayB (1).
XAXS (?) d {, i}
The XAXS command sets (queries) the X Axis Scale Type of display d. The
parameter i selects Linear (0) or Logarithmic (1).
This command is valid only when display d is an FFT measurement with
frequency as the X axis. FFT measurements with a time X axis (Time, WinTime,
Capture or User Time Function) always have a linear scale.
RPMF (?) {i}
The RPMF command sets (queries) the RPM frequency scaling for both
displays. The parameter i selects Hz (0) or RPM (1) as the units for frequency.
GRID (?) d {, i}
The GRID command sets (queries) the Grid On/Off Mode of display d. The
parameter i selects Off (0) or On (1).
GDIV (?) d {, i}
The GDIV command sets (queries) the number of Grid Divisions of display d.
The parameter i selects 8 (0), 10 (1), 12 (2) or 15 (3).
Changing the Grid Divisions changes the vertical scaling (Y/div) and horizontal
scaling (X/div) (Nyquist and Nichols views).
GPOL (?) d {, i}
The GPOL command sets (queries) the Nyquist Grid type for display d. The
parameter i selects Rectangular Grid (0) or Polar Grid (1).