Exceedance
Statistics
1-61
5. Press [Average]
Select the Average menu. Note that this menu is
changed in Octave group.
We will leave the averaging at its default, 100 ms
exponential time.
Press <Power Bin>
Change the Power Bin.
Use the knob to select (L) and press [Enter].
Choose L (L
eq
) as the sound level bin. L
eq
is a
standard broad band sound measurement. The
result is displayed as the last bin in the display and
is labeled ‘L’.
Press [Auto Scale A]
Scale DisplayA to show the entire range of the data.
6. Press [Waterfall]
Select the Waterfall menu. L
n
is calculated from
measurements stored in the waterfall buffer.
Press
<Storage>
Select waterfall Storage.
Use the knob to select (One Shot) and press
[Enter].
Choose One Shot to fill the waterfall buffer once
and stop.
Press <Total Count>
Change the number of measurements to store in the
waterfall buffer.
Press [1] [0] [0] [Enter]
Enter the Total Count for both displays (100).
The Storage Interval is 100 ms. A measurement
snapshot is added to the waterfall buffer every
Storage Interval so 100 measurements will take 10 s
to complete.
Press <Save Option>
The SR785 has two waterfall storage options for
saving measurements. (All) Means that all possible
measurements for the currently selected
measruement group will be stored in the waterfall
buffer. This option saves more measurements but
each trace takes more space in the waterfall buffer,
so fewer traces will be available.
(Active Meas Only) means that only the currently
selected measurement will be saved in the waterfall
buffer.
Use the knob to select (Active Meas Only) and
press [Enter]
To calculate exceedence statistics, we must use the
Active Measurement Only save option.