5-90 Exceedance Statistics Commands
SR785 Dynamic Signal Analyzer
Exceedance Statistics Commands
ESTR (?) {i}
The ESTR command sets (queries) the Exceedance Start Index (for both
displays). The parameter i is a waterfall index [0..].
This command is not valid when the measurement group is swept-sine.
ESTP (?) {i}
The ESTP command sets (queries) the Exceedance Stop Index (for both
displays). The parameter i is a waterfall index [0..]. Remember, the records are
numbered starting from 0. The last record is the number of stored records minus
one. The exceedance calculation can not proceed if the stop index is greater than
the number of records stored in the waterfall buffer minus one.
This command is not valid when the measurement group is swept-sine.
EPCT (?) {i}
The EPCT command sets (queries) the Exceedance Centile (for both displays).
The parameter i is a centile from 1 to 99.
This command is not valid when the measurement group is swept-sine.
EXCE d, i
The EXCE command starts the exceedance centile calculation for display d. The
parameter i selects a Trace [1..5] to store the result.
Display d must be paused or done with waterfall storage on and records stored in
the waterfall buffer using the active measurement only save option. Waterfall
display is not required to be on.
The result is stored in a data trace and has the same measurement type as the
waterfall measurements.
This command is not valid when the measurement group is swept-sine.