Time/Histogram Frequency Menu 4-23
SR785 Dynamic Signal Analyzer
Time/Histogram Frequency Menu
When the Measurement Group is Time/Histogram, this menu sets parameters related to
the resolution and number of points in the histograms.These parameters govern the
measurements on both displays. See ‘Time/Histogram Analysis’ in Chapter 2 for a
discussion about time/histogram measurement fundamentals.
Sample Time
Sets the sampling interval for time records and histograms in the Time/Histogram
measurement group. The allowed sampling times depend on the setting of the base
frequency softkey. If the base frequency is 102.4 kHz the allowed sampling times range
from 3.81
µ
s to 2s in powers of 2. If the base frequency is 100.0 kHz the allowed
sampling times range from 3.91
µ
s to 2.048s in powers of 2.
Command: SPAN (?) d {, i}
Histogram Length
Sets the number of points in each histogram. The number of points can be specified in
samples, records, (each record is 1024 samples), or in seconds.When the specified
number of samples, records, or seconds, has been reached, the histogram will update.
If repeat is on the SR785 will begin accumulating a new histogram at the end of the prior
histogram.
Command: HLEN (?) d {, f}
Bins
Sets the number of histogram 'bins' along the x-axis. Values from 4 to 1024 in power of 2
increments are allowed. The x-axis display for histogram measurements ranges from -
1.25 times the input range to +1.25 times the input range. The resolution of the histogram
is the therefore 2.5 times the input range divided by the number of bins.
Command: HBIN (?) d {, i}