4-114 Average Menus
SR785 Dynamic Signal Analyzer
can be changed so as to preview any other measurement in the current measurement
group. Subsequent time records will return to the newly selected measurement rather
than the time record for preview.
Command: PAVO (?) d {, i}
Preview Time
Select the Preview Time for both displays [0.5 s .. 1 ks].
When Average Preview is Timed, the input time record is automatically accepted after
the Preview Time (unless rejected first).
Command: PAVT (?) d {, x}
Accept
Accept the displayed preview time record and add the measurement to the average.
Average Preview allows each individual measurement to be accepted (added to the
average) or rejected (not added to the average) based upon the input time records. This is
useful for rejecting bad time records from corrupting an averaged measurement.
Command: PAVA
Reject
Reject the displayed preview time record and do not add the measurement to the average.
Average Preview allows each individual measurement to be accepted (added to the
average) or rejected (not added to the average) based upon the input time records. This is
useful for rejecting bad time records from corrupting an averaged measurement.
Command: PAVR