5-62 Input
Commands
SR785 Dynamic Signal Analyzer
Transducer Parameter Commands
EU1M (?) d {,i}
The EU1M command sets (queries) the Ch1 Engineering Units Mode for display
d. The parameter i selects Off (0) or On (1).
The set command requires d = 2 (both displays).
EU1L (?) d {,i}
The EU1L command sets (queries) the Ch1 Engineering Units Label for display
d The set command requires d = 2 (both displays).
The parameter i selects the unit label.
i Label
i Label
0
m/s
2
8
kg
1 m/s
9 lbs
2
m
10
N
3
in/s
2
11
dyne
4
in/s
12
Pas
5
in
13
bar
6
mil
14
USER
7
g
EU1V (?) d {,x}
The EU1V command sets (queries) the Ch1 Engineering Units per Volt scale for
display d The set command requires d = 2 (both displays). The parameter x is
real number of EU/Volt (EU’s per Volt).
EU1U (?) d {,s}
The EU1U command sets (queries) the Ch1 User Label for display d The set
command requires d = 2 (both displays). The string s is the user label.
TD1C (?) d {,i}
The TD2C command sets (queries) the Ch1 Transducer Conversion. for display
d. The set command requires d = 2 (both displays). The parameter i selects the
units to be converted to:
i Label
0
m/s
2
1
m/s
2
m
3
in/s
2
4
in/s
5
in
6
mil
7
g
Содержание SR785