FFT Frequency Menu 4-11
SR785 Dynamic Signal Analyzer
This softkey also anchors the End frequency. Adjusting the FFT Span will leave the End
frequency unchanged. Enlarging the frequency Span may change the End frequency if
the new measurement would extend below 0 Hz or above 102.4 (100.0) kHz.
If the measurement group is Correlation, then the End frequency is always equal to the
Span (baseband).
The two displays can have different End Frequencies if the Analyzer Configuration is set
to Independent Channels. In this configuration, no two channel measurements are
allowed (frequency response, cross-spectrum, etc.), but the entry field can be linked or
unlinked using the [Link] key. If Analyzer Configuration is set to Dual Channel they
field is automatically linked to both displays.
Command: FEND (?) d {, f}