Methods, Test Panels, Article Management
Phadia 100 - User Manual
P a g e 9 . 8
12-3501-20/12
** This menu can not be reached for the default methods.
5.2.3.1.2.2.Identity**
Type the identity of the curve control. Confirm with
enter
.
5.2.3.1.2.3.Test name**
Type the Test name of the ImmunoCAP, maximum 5 characters, that will be used
together with the curve control, confirm with
enter
.
5.2.3.1.2.4.Replicate**
Type the number of replicates (normally 1), confirm with
enter
.
5.2.3.1.2.5.Concentration**
Type the expected concentration, confirm with
enter
.
5.2.3.1.2.6.Expected s
Expected standard deviation. Type the figure, confirm with
enter
.
The inner and outer limits for the curve control are expressed as a factor that will
be multiplied with the expected standard deviation.
5.2.3.1.2.7.Limit s Inner
Inner limit of the concentration of the curve control. This factor will be multiplied
with the Expected s. The limit is for both above and below expected concentra-
tion.
Any change will be printed on the laboratory report.
5.2.3.1.2.8.Limit s Outer
Outer limit of the concentration of the curve control. This factor will be
multiplied with the Expected s. The limit is for both above and below expected
concentration.
Any change will be printed on the laboratory report.
5.2.3.1.2.9.Log start date
Type the log start date in the displayed format (year-month-day). Confirm this
step with
enter
.
Any change will be printed on the laboratory report.
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...