Assay Run Editing Functions
12-3501-20/12
Page 4.25
3
4
5
6
4
In the Edit QC mode it is possible to change status, identity, test name, replicate,
response, expected concetration, standard deviation, limit standard deviation,
range minimum and maximum, log start date and lot number.
To reach this function, you first have to login with the laboratory password (the
password at delivery is “unicap”).
1.
In Prepare Run mode
press
5
.
2.
In Method Setup mode
press
yes
.
3.
Press
2
.
4.
In Edit Method mode
press
yes
.
5.
Select the method for
which you want to define a
Quality Control by pressing
no
until the desired method is dis-
played. Then press
yes
.
6.
Press
9
.
7.
In Quality control mode
press
yes
.
8.
Press
2
.
9.
In Edit QC mode press
yes
.
5.Method Setup
2.Edit Method?
5.Method Setup
1.Add Method?
5.2.Edit Method
Method: ImmunoCAP Specific IgE 0-100?
5.2.Edit Method
Please wait...
5.2.Edit Method
[sIgE]
1.General parameters?
5.2.Edit Method
[sIgE]
9.Quality control?
Phadia 100 Ver.3.0
14:10:17
5.Method Setup?
Phadia 100 Ver.3.0
14:10:17
1.Prepare Run?
5.2.9.Quality control
[sIgE]
1.Add QC?
5.2.9.Quality control
[sIgE]
2.Edit QC?
Edit QC
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...