Phadia 100 - User Manual
Error Handling
12-3501-20/12
Page 12.45
12
Corrective actions:
No corrective actions proposed.
If the error remains contact your local Phadia representative.
CC-1 HIGH* and CC-2 HIGH*
What can cause the problem:
This error might be caused by improper placement of the instrument, but the
results might be acceptable. For
EliA
it can be a contamination problem (mostly
contamination of conjugate or CC vial whith sample) or wrong Lot Specific
Code.
Corrective actions:
It is recommended though, that the reliability of the results be judged using QC
results. For additional information see Explanation of Actions, number
7, 22, 26,
30, 32.
If the error remains contact your local Phadia representative.
CC-1 HIGH* and CC-2 LOW*
What can cause the problem:
This error is very uncommon and is probably caused by more than one error.
Corrective actions:
No corrective actions proposed, except for
EliA
methods
.
Follow instructions in Action List in order to identify or eliminate the errors. For
additional information see Explanation of Actions, number
7, 18, 24.
If the error remains contact your local Phadia representative.
CC-1 HIGH* and CC-2 HIGH
What can cause the problem:
For
EliA
it can be a contamination problem (mostly contamination of conjugate
or CC vial whith sample) or wrong Lot Specific Code. For other methods this
error is very uncommon and is probably caused by more than one error.
Corrective actions:
No corrective actions proposed, except for
EliA
methods.
Follow instructions in Action List in order to identify or eliminate the errors. For
additional information see Explanation of Actions, number
7, 18, 22, 24, 25,
26, 27, 30, 32.
If the error remains contact your local Phadia representative.
CC-1 HIGH* and CC-2 LOW
What can cause the problem:
This error is very uncommon and is probably caused by more than one error.
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...