Phadia 100 - User Manual
Maintenance
12-3501-20/12
Page 11.13
2
5
6
11
Unscheduled Maintenance
The following points are included in the unscheduled maintenance:
• Actions before Shutdown and Startup
• Selftest
• Blank Run
• FluoroC Run
• Loading of Printer Paper
• Changing of Fuses
• System Rinse
Actions Before Shutdown and Startup
For Phadia 100 Ver.3.0, use the menu
6.13.Shutdown
and follow the instruc-
tions displayed to safely shut down the instrument.
Before switching Phadia 100 off for longer than a month a daily maintenance has
to be processed twice. The first rinse has to run with purified water, followed by
a run without any liquid at all.
Wipe off any remaining liquid from the Processing Chamber.
Before starting up again, perform a monthly maintenance and a selftest.
Selftest
The selftest checks various functions of the instrument.
To run a selftest you need Development Solution and Stop Solution. Make sure
that the Washing Solution and Rinse Solution Bottles are filled and connected to
the instrument.
1.
In Prepare Run mode se-
lect the number
6
key.
Phadia 100 Ver.3.0
07:55:06
1.Prepare Run?
WARNING! Washing Solution may cause
sensitization by skin contact. Wear protective
gloves!
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...