Phadia 100 - User Manual
Error Handling
12-3501-20/12
Page 12.15
12
Example of an Error List and an Action List printed at the end of a Laboratory Report:
A question mark (?) is printed on the Laboratory Report indicating that a
Processing Error has occured for the calibrator, curve control, Quality Control
or sample.
Example of a question mark printed on the Laboratory Report:
If a Processing Error has occured, a Distribution List is automatically printed
after the Laboratory Report. By comparing the information underneath PROC-
ESS ERRORS with the information on the Distribution List you can identify the
affected position in the Sample Carousel and the component involved in the
error. With this information it might be easier to identify and eliminate the cause
of the error. Note that one sample might be found in several positions in the
ImmunoCAP Carousel.
SAMPLES
-TEST---RESP---CONC---CLASS---QUOT--
32418
ImmunoCAP Specific IgE (kUA/l)
t3
1591?
7.09
3
SEE ACTION LIST FOR CORRECTIVE ACTIONS
The question mark indicates
which result the processing error
has affected
DISTRIBUTION LIST
Run number:
1
Run created:
090507 14:29
Process started:
090507 15:23
Process ended:
090507 17:51
Run origin:
Instrument
SAMPLE CAROUSEL
A sIgG
Conjugate
E
Development solution
G IgA/IgG diluent Diluent
H
Stop solution
01 sIgG CAL-0.02
Cal
02 sIgG CAL-0.04
Cal
03 sIgG CAL-0.1
Cal
04 sIgG CAL-0.3
Cal
05 sIgG CAL-1.0
Cal
06 sIgG CAL-2.0
Cal
Cont...
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...