Maintenance
Phadia 100 - User Manual
P a g e 1 1 . 1 4
12-3501-20/12
2.
Press the
yes
key.
3.
In Install/Service mode
select the number
9
key.
4.
Press the
yes
key.
5.
In Maintenance mode se-
lect the number
6
key.
6.
Press the
yes
key.
7.
Press the
enter
key to start
the selftest.
8.
Load the Development
Solution vial in position E. Press
the
enter
key to confirm this
step.
9.
Load the Stop Solution
vial in the given position. Press
the
enter
key to confirm this
step.
10.
Press the
enter
key to start
the process.
After the selftest is finished the instrument will automatically print a Blank
Report where you can see the result of the selftest. If you have received a section
on the printout with remarks you have a Blank Error. An action list will then be
printed at the end of the report, with instructions on how to eliminate the error.
For more information on how to eliminate the error, see chapter 12, Error
Handling.
6.9.Maintenance
1.Daily Maintenance?
6.9.Maintenance
6.Run Selftest?
6.9.6.Run Selftest
PRESS ENTER TO START A BLANK CHECK
6.9.6.Run Selftest
LOAD DEV SOL IN POS E
6.9.6.Run Selftest
PRESS ENTER TO START PROCESS
6.9.6.Run Selftest
LOAD STOP SOL IN POS H
6.Install/Service
9.Maintenance?
6.Install/Service
1.Login/Logout?
Phadia 100 Ver.3.0
07:55:06
6.Install/Service?
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...