Phadia 100 - User Manual
Error Handling
12-3501-20/12
Page 12.11
12
Example of a blank report:
The blank check errors are specified underneath Remarks on the Blank Report.
At the end of the Blank Report you find the action list with instructions on how
to eliminate the errors.
Four different blank check errors can occur:
REAGENT BLANK IS TOO HIGH
If the Reagent blank is >187.5 RU this error message will occur.
What can cause the problem:
This error might be caused by contamination of the Development Solution or the
Elution Wells Disc might need cleaning.
0354, 090507 10:12, 2.3, 032
BLANK REPORT
Run date, time:
090507 10:00
Source:
Before Assay
Reagent Blank:
13.5 RU
Rinse Blank:
11.7 RU
Dark Blank:
9.4 RU
REMARKS
THE DIFFERENCE BETWEEN REAGENT BLANK
AND RINSE BLANK IS TOO LOW
SEE ACTION LIST FOR CORRECTIVE ACTIONS
ACTION LIST
The Cuvette is probably broken
Run Selftest, menu 6.9.6.
If Blank Error remain - Contact Phadia representative
The Blank Check errors are printed
here
Action List with instructions on
how to eliminate the error
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...