Assay Run Editing Functions
12-3501-20/12
Page 4.41
3
4
5
6
4
Remove Single ImmunoCAP/EliA Well
2.1.Abort Prepared Run
1.Delete all Request Data?
2.1.Abort Prepared Run
2.Remove Single ImmunoCAP/Elia Well?
This is used in case the wrong ImmunoCAP/EliA Well was loaded, and this was
noticed before starting the process.
1.
In the Abort Prepared Run
mode press
no
.
2.
Press
yes
.
Warnings that may prevent entering
PROCESS IS RUNNING
You cannot remove any ImmunoCAP/EliA Well when an assay run is being
processed.
3.
Press
no
to move the
ImmunoCAP carousel to the
next position. Press
no
and
shift
at the same time to move
the carousel to previous posi-
tion.
You have to remove the ImmunoCAP/EliA Well from the displayed position
manually. Press
lid open
to open the processing lid.
Find the correct ImmunoCAP/EliA Well and remove it manually.
Close the lid. Now you can press
enter
.
There will be a message telling you that the ImmunoCAP/EliA Well is removed.
ImmunoCAP/EliA Well will get the status not loaded in the prepared run.
To exit press
mode
.
NOTE!
If there is no ImmunoCAP/EliA Well with status loaded in the prepared
run there will only be a position displayed and
enter
has no function.
Messages that can occur
POSITIONING...
The ImmunoCAP Carousel is moving to a new position.
2.1.2.Remove Single ImmunoCAP/EliA Well
23: sIgE t17 PRESS ENTER IF REMOVED
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...