Maintenance
Phadia 100 - User Manual
P a g e 1 1 . 1 2
12-3501-20/12
Replace Drainage
Filter
At the bottom of the Processing Chamber
there is a drainage filter (12-3500-75) that
prevents particles from entering valves. It
is advisable to replace it every six months.
Press
lid open
and open the Processing
Lid.
Pull out the drainage filter from the Pro-
cessing Chamber using a pair of forceps
or pliers. Replace it with a new drainage
filter. In case waste does not disappear
from the processing chamber the drain-
age filter can be changed more often.