Phadia 100 - User Manual
Routine Assay Run
12-3501-20/12
Page 2.47
2
3
4
5
6
2
13
LOT SPECIFIC CODE NOT EQUAL
Present Lot Specific Code is not the same as the code used for the calibrators.
15.
When all ImmunoCAP/
EliA Wells are loaded, remove
the carrier. A message tells you
that everything is loaded. Press
enter
to confirm this step.
16.
Press
yes
to be able to
start processing the assay run.
If you press
no
you will return
to the Load Reagents mode.
17.
Normally this step does
not occur. Only the service
technican can activate a vial
check. Press
enter
to start it.
A message occurs that the pro-
cess has started.
If the vial check is ok, continue with step 19, otherwise with step 18.
Warnings that may prevent entering
LOAD IS NOT COMPLETED
Before the run can be started, reagents, samples and ImmunoCAP/EliA Well
must be loaded.
PRESSURE TOO LOW (1200 hPa)
The system pressure is too low. The system pressure has to be at a certain level
in order for the system to produce reliable results. 1200 is the actual pressure
in hPa.
WAITING FOR TEMPERATURE
Before the run can be started, incubation temperature has to reach the correct
level. The run will automatically start when the right temperature is reached.
POSITIONING...
This message will appear when the Sample Carousel is moving to a new position.
DISPENSING...
This message will appear when the instrument is dispensing ImmunoCAP/EliA
Well.
1.3.4.Start Process
PRESS ENTER TO START A VIAL CHECK
1.3.Load and Start
4.Start Process?
1.3.3.Load ImmunoCAP/EliA Well
ALL LOADED. REMOVE CARRIER. PRESS ENTER
1.3.4.Start Process
PROCESS STARTED
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...