Phadia 100 - User Manual
Error Handling
12-3501-20/12
Page 12.3
12
When the system is started, a number of checks are made to ensure that
everything is running correctly. If an abnormality is found, it results in a Start
up error or a Start up warning.
Start Up Errors
Errors are normally issued for serious malfunctions when there is no meaning of
continuing the start up. If an error is displayed, the system halts and the operator
is instructed to call for service. Start by switching
Phadia 100 off, wait 10 seconds and switch the instrument on again.
If the error returns make a note of the error code and contact your local Phadia
representative.
Example of a start up error:
Start Up Warnings
If the malfunction is of a less serious type, a warning is issued. The operator is
instructed to note the warning and then press any key. Example of a start up
warning:
When a key has been pressed, the startup continues.
If the warning returns contact your local Phadia representative.
Start Up Errors and Start Up Warnings
STARTUP WARNING #2301
Note warning, then press Any key...
STARTUP ERROR #1201
Startup was aborted. Call service...
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...