Instrument Specification and Installation
Phadia 100 - User Manual
P a g e 1 . 1 0
12-3501-20/12
Working Conditions
Ambient temperature: 18ºC to 32ºC
Ambient air humidity: 10% to 85% RH
Ambient air pressure: 860 hPa to 1060 hPa
Ambient temperature change: 2ºC/hour
(max.)
Heat Contribution
310 W (average) 1058 Btu/h
Storage Conditions
Ambient temperature: -20ºC to +70ºC
Ambient air humidity: 10% to 95% RH
Ambient Sound Level
57 dBA
Warm-Up Time
For normal assay: 1 hour
Interface
ImmunoCAP 100 equipped with USB ports:
Two USB 2.0 type A ports are located at the
upper right of the instrument and two USB 2.0
type A ports are located at the instrument rear.
The USB ports can be used to connect USB
memorys or external USB harddrives.
One TCP/IP connector is located at the upper
right of the instrument rear. The TCP/IP
connector can be used for connection to IDM
(Phadia Information Data Manager).
One serial port (COM 2) is located at the upper
left of the instrument rear. (COM 1 is used
internally in the instrument.) This connection
will handle serial data communication in
accordance with the RS 232 C standard. The
COM 2 port can be used to connect a bar code
reader.
Pin characteristics (pin no., name, output/ input,
maximum voltage).
Pin 3 Tx
output
±12 V
Pin 2 Rx
input
±12 V
Pin 7 RTS
output
±12 V
Pin 8 CTS
input
±12 V
Pin 6 DSR
input
±12 V
Pin 4 DTR
output
±12 V
Pin 5 GND
signal ground
Pin 1 DCD
input
±12 V
Pin 9 RI
input
±12 V
The serial port must be connected only to
external equipment meeting the appropriate
EMC and electrical safety standards
(i.e., IEC 1010-1, IEC 601-1, IEC 950,
EN 61010-1, EN 60601-1, EN 60950 or
equivalent UL-standard)
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...