Maintenance
Phadia 100 - User Manual
P a g e 1 1 . 2 0
12-3501-20/12
Replace Fuses
Extra primary and secondary fuses are included in the accessory box.
NOTE!
Replace only with fuses of specified type and rating.
Primary Fuses
1.
Disconnect the mains
power cord.
2.
Open the fuse/voltage se-
lector cover using a screwdriver.
Insert the screwdriver at the
top center and apply pressure.
3.
Take out the fuse holders
from the instrument and re-
move the old fuses.
4.
Place new fuses in the fuse
holders and install them in both
the left- and right-hand posi-
tion in the instrument.
5.
Close the cover.
6.
Connect the mains power
cord.
NOTE! Phadia 100 must only be connected to a
grounded power outlet (protective ground).
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...