Phadia 100 - User Manual
Routine Assay Run
12-3501-20/12
Page 2.39
2
3
4
5
6
2
13
1234 (sIgE)Test:_ [44]
t3, w1, g6, f1
1234 (sIgE)Test:f1_
[45]
t3, w1, g6
1234 (sIgE)Test:_ [48]
1.1.Enter Request
[48]
Enter Sample ID:_
3.
Type the sample ID (up to
16 characters) and press
enter
.
If you use a barcode reader, the
sample ID is entered when you
read the barcode.
To add a calibration curve type
cal
instead.
To add Quality Control type
qc
(if status MANUAL is chosen in the method).
4.
Enter the test names or
test panel names by typing them
one at a time and press
enter
to
confirm each test name.
If you use a barcode reader,
each test is entered when you
read the barcode.
To change method press
alt
and choose the method. Then select the tests you
want for this method.
If you want to change replicates or dilution factor, please see chapter 4, Assay
Run Editing Functions.
5.
To save the entered tests
for the sample identity, press
enter
again.
NOTE!
If you press the
mode
key before
enter
, you will lose the information
entered to the sample identity.
6.
Now you can enter the
next sample identity or leave
this mode.
If you want to return to Enter Request mode, press
mode
or
enter
.
When you entered all the samples for the request you may continue processing
the assay run. Then go to
1.3.Load and Start.
7.
When no positions are left
in the instrument, the next en-
tered test will induce a ques-
tion.
You can skip the sample identity or just the test that cannot be positioned. Press
yes
if you want to skip the sample ID. All the tests entered for this sample ID
will be deleted.
Press
no
if you do not want to skip the sample ID.
1.1.Enter Request [44]
Enter Sample ID:_
1235 (sIgE)Test:f3
[0]
NO TEST POS. LEFT. Skip Sample ID?
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...