Phadia 100 - User Manual
Assay Run Quality Control
12-3501-20/12
Page 6.5
6
Edit Log
4.Quality Control
1.Print Log?
4.Quality Control
2.Edit Log?
4.2.Edit Log
Method: ImmunoCAP specific IgE 0-100?
You can inactivate and activate points in the quality control and curve control
logs.
1.
In Prepare Run mode
press
4
.
2.
Press
yes
.
3.
In Print Log press
2
.
4.
Press
yes
.
5.
Press
no
until the method
you want appears, then press
yes
.
Warnings that may prevent entering
CANNOT SELECT THIS METHOD
The selected method cannot be used for this operation.
Edit Quality Control-Log
You can set the status to active or inactive for a specific QC control point. If you
change a point to the status inactive, this point will be marked inactive when the
log is printed and it will not be used in the calculation of mean and variation.
1.
In Edit Log mode press
no
until the method you want
appears, then press
yes
.
2.
Press
yes
.
4.2.Edit Log
Method: ImmunoCAP specific IgE 0-100?
4.2.Edit Log
[sIgE]
1.Quality control-log?
Phadia 100 Ver.3.0
14:10:17
1.Prepare Run?
Phadia 100 Ver.3.0
14:10:17
4.Quality Control?
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...