Error Handling
Phadia 100 - User Manual
P a g e 1 2 . 5 2
12-3501-20/12
Example from a Laboratory Report:
LOW RU
Check the Reagent Blank under Remarks on the Laboratory Report. If the
Reagent Blank is higher than expected (>20 RU for ImmunoCAP and >35 RU for
EliA Wells), the error might be caused by insufficiently cleaned Elution Wells
Disc or that Stop Solution has not been pipetted at the end of the run.
If the reagent blank is higher than expected, perform a complete Daily
Maintenance.
Open the Processing Lid and check for any ImmunoCAP or EliA Wells blocking
the movement of the ImmunoCAP Carousel. If there is a misplaced ImmunoCAP
or EliA Well remove it. If the ImmunoCAP or EliA Wells are still in their inner
positions there is a mechanical malfunction.
If you have received LOW RU for all positions, load the assay run again by using
1.5.Load Old Run
and rerun the samples.
If you have received LOW RU for some positions, rerun the results that have not
been reported.
SAMPLES
-TEST---RESP---CONC---CLASS---QUOT--
32418
ImmunoCAP Specific IgE (kUA/l)
t3
LOW RU
--.--
----
33279
ImmunoCAP Specific IgE (kUA/l)
w6
1591
7.09
3
SEE ACTION LIST FOR CORRECTIVE ACTIONS
REMARKS
REAGENT BLANK VALUE (14)
SEE ACTION LIST FOR CORRECTIVE ACTIONS
ACTION LIST
See User Manual, Index: Error handling LOW RU
Check/remove misplaced ImmunoCAP/Wells in the ImmunoCAP Carousel
Check if Reagent Blank has increased. See Laboratory Report, REMARKS
Rerun the results that have not been reported
LOW RU is printed instead of the
response. No results are pre-
sented
The Reagent Blank value is printed
here
Action List with instructions on
how to eliminate the error
0354, 090507 10:12, 2.3, 032
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...