Error Handling
Phadia 100 - User Manual
P a g e 1 2 . 1 0
12-3501-20/12
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors.
Ensure that Washing Solution and Rinse Solution are properly prepared. The
Washing Solution Concentration is of big importance for several tests. Always
prepare according to Directions for Use. Always use purified water when
preparing Washing Solution and Rinse Solution.
Check that Rinse Solution is loaded in Rinse bottle and Washing Solution is
loaded in Washing Solution Bottle. Shake both Washing and Rinse Solution
bottle, the bottle with Washing Solution will foam.
If the error remains contact your local Phadia representative.
INTERNAL ERROR
There is an internal system error. It is not possible to start the assay run.
Contact your local Phadia representative.
DAILY MAINTENANCE NOT PERFORMED
The daily maintenance has not been performed within 24 hours after the last
completed run. It is important to perform daily maintenance in order to keep the
instrument in good condition.
Please perform daily maintenance every day, see chapter 11, Maintenance.
WEEKLY MAINTENANCE NOT PERFORMED
The weekly maintenance is required. It is important to perform weekly
maintenance in order to keep the instrument in good condition.
Please perform weekly maintenance once a week, see chapter 11, Maintenance.
MONTHLY MAINTENANCE NOT PERFORMED
The monthly maintenance is required. It is important to perform monthly
maintenance in order to keep the instrument in good condition.
Please perform monthly maintenance once a month, see chapter 11, Mainte-
nance.
BLANK CHECK ERROR
Blank Errors might be detected when the instrument performs a Blank Check. The
Blank Check is used to test the reagent blank, which is subtracted from the results
obtained in the assay run. The Blank Check also tests the quality of the Rinse Solution
and the function of the cuvette and fluorometer. When Blank Check Error is
displayed a blank report will automatically be printed.
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...