Routine Assay Run
Phadia 100 - User Manual
P a g e 2 . 4 6
12-3501-20/12
11.
Press
yes
to start loading
ImmunoCAP/EliA Wells.
12.
Press
enter
to start a
blank check.
If a blank error occurs, se chapter 12, Error Handling.
Messages that can occur
WASH/RINSE CHECK NOT OK
For corrective actions see Action list.
WASH/RINSE HAVE BEEN MIXED
For corrective actions see Action list.
13.
After the blank check is
finished start loading
ImmunoCAP/EliA Wells.
Place the required ImmunoCAP/EliA Well carrier in position. Assure yourself
that you are loading the correct ImmunoCAP/EliA Well. The instrument will
distribute the ImmunoCAP/EliA Well to the correct position.
14.
Type the Lot Specific
Code that is found on the label
on the EliA Well foil bag and
on the label on the EliA Well
Carrier. Press
enter
to confirm
the Lot Specific Code.
Messages that can occur
WRONG LOT SPECIFIC CODE
Wrong code entered.
USE DEFAULT LOT SPECIFIC CODE?
By pressing
shift
and
enter
or by pressing
enter
without entering a code, this
question is shown.
LOT SPECIFIC CODE CORRECT?
If correct press
yes
. If no, you have to select a tube with the same code as the
calibrators. Press
enter
to confirm the Lot Specific Code.
1.3.3.Load ImmunoCAP/EliA Well
LOAD 12 sIgE a_IgE
1.3.3.Load EliA Well Code:_
LOAD 12 El-G Gcal
1.3.3.Load ImmunoCAP/EliA Well
PRESS ENTER TO START A BLANK CHECK
1.3.Load and Start
3.Load ImmunoCAP/EliA Well?
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...