FluoroC %CV: NOT OK
If you get NOT OK after FluoroC %CV on the FluoroC Run Report the
instrument might have run out of FluoroC solution during the FluoroC run or
there might be a problem with the pipette.
Make sure that the tubings from the Washing Solution and Rinse Solution
Bottles are not twisted and that the tubings are properly connected to the bottles.
Perform a complete cleaning procedure according to Monthly Maintenance. For
further instructions see chapter 11, Maintenance.
Rerun FluoroC using new room temperated vials of FluoroC and Stop Solution.
If the error remains contact your local Phadia representative.
Deviation from target:
[±4-10%]
NOT OK
If you get NOT OK after Deviation from Target on the FluoroC Run Report and
the deviation is between 4% and 10% or -4% and -10%, rerun FluoroC using
new room temperated vials of FluoroC and Stop Solution.
If the error remains contact your local Phadia representative.
Deviation from target:
[>+10% or <-10%]
NOT OK
If you get NOT OK after Deviation from Target on the FluoroC Run Report and
the deviation is more than 10% or less than -10%, the instrument might be out
of Stop Solution.
Make sure that the tubings from the Washing Solution and Rinse Solution Bottles
are not twisted and that the tubings are properly connected to the bottles.
Run
6.9.6.Selftest
. For further instructions see chapter 11, Maintenance.
Clean and dry Elution Wells Disc.
Rerun FluoroC using new room temperated vials of FluoroC and Stop Solution.
If the error remains contact your local Phadia representative.
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...