Maintenance
Phadia 100 - User Manual
P a g e 1 1 . 1 8
12-3501-20/12
6.9.3.1.FluoroC Run
LOAD 180 ul STOP SOL IN POS H
6.9.3.1.FluoroC Run
PRESS ENTER TO START PROCESS
If the displayed target value is to be used, press the
enter
key.
If a new target value is to be entered, type the value and press the
enter
key.
Confirm the new target value by typing it again and pressing the
enter
key.
9.
Load the Stop Solution
vial in the given position. Con-
firm this step by pressing the
enter
key.
10.
To start the process press
the
enter
key.
After the FluoroC run is finished the instrument will automatically print a
FluoroC Run Report. On the report you can see the result of the FluoroC run.
If you have received a NOT OK on the printout for FluoroC %CV, Rinse Blank
voltage, or Deviation from target, you have a FluoroC error. Follow the short
instructions on the action list on the printout to eliminate the error.
For further information on how to eliminate the error, see chapter 12, Error
Handling.
Example of a FluoroC Run Report:
FLUOROC RUN
Run date, time:
090513 11:53
Measurement status: OK
Target value:
DW (12500 RU)
RESULTS------------Value--Acceptance
FluoroC voltage 1:
3923.0 mV
FluoroC voltage 2:
3932.3 mV
FluoroC voltage 3:
3946.1 mV
FluoroC Mean voltage:
3933.8 mV
FluoroC %CV:
0.30 %
OK
Rinse Blank voltage:
4.0 mV
OK
Final FluoroC voltage: 3929.8 mV
Final FluoroC RU:
12280.5 RU
Deviation from target: -1.8 %
OK
0354, 090513 12:12, 2.3, 032
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...