Error Handling
Phadia 100 - User Manual
P a g e 1 2 . 6
12-3501-20/12
These errors may occur during the preparation of an assay run. An Action List
with short instructions will be printed every time a message is displayed.
Example of a preparation error:
The following Action List will be generated:
Follow the instructions and perform the necessary steps to eliminate the error.
If needed you will find more detailed information in this chapter. When the error
is eliminated you will be able to continue assay preparation.
SAMPLE CAROUSEL COVER IS NOT PRESENT
What can cause the problem:
The black device on the left side of the Sample Carousel has not been pushed in.
Indicating that that Sample Carousel Cover is not present/in position.
Corrective actions:
Put the Sample Carousel Cover in position. Will minimize evaporation from
samples and reagents and also to protect them from contamination.
PROCESSING LID IS NOT CLOSED
What can cause the problem:
The sensor in the Process Chamber gets information that the Processing Lid is
not closed. If the Processing Lid is not closed several functions in the instrument
will not work.
Corrective actions:
Close the Processing Lid properly and make sure nothing is obstructing the lid.
Preparation Errors
LAMP VOLTAGE TOO LOW
ACTION LIST
Check/replace fuse F904 and F906
If error remain - contact Phadia representative
0354, 090507 10:12, 2.3, 032
Action List with instructions on
how to eliminate the error
The Preparation Error is printed
here
1.3.Load and Start
LAMP VOLTAGE TOO LOW
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...