Assay Run Editing Functions
Phadia 100 - User Manual
P a g e 4 . 2 8
12-3501-20/12
3.
Delete or type over the
old number of replicates. Press
1
,
2
,
3
,
4
or
5
depending on
how many replicates you want
to run. Press
enter
to confirm
this step.
Messages that can occur
Replicate: TOO HIGH!
The number of replicates is too high. Please, choose a number between 1 and 5.
Replicate: TOO LOW!
The number of replicates is too low. Please, choose a number between 1 and 5.
Coefficient of variance limit response
The following step describes how you can change the maximum allowed % CV.
1.
In Edit QC mode press
5
.
2.
Press
yes
.
3.
Delete or type over the
old maximum allowed % CV.
Type the new maximum al-
lowed % CV and press
enter
to confirm this step.
Expected concentration
The following step describes how you can change the expected mean value. It
is also possible to deactivate this function.
After adding a Quality Control it is recommended to run at least 20 assay runs
to get the expected mean value in the Quality Control log.
You find the expected mean value at the end of the Quality Control log.
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
1.Status?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
5.%CV limit resp?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
%CV limit resp: 0.1
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
Replicate: 1
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...