Assay Run Editing Functions
Phadia 100 - User Manual
P a g e 4 . 3 2
12-3501-20/12
2.
Press
yes
.
3.
Delete or type over the
old highest allowed concentra-
tion. Type the new highest al-
lowed concentration or type 0
(zero) to deactivate the func-
tion. Press
enter
to confirm
this step.
If you are using ImmunoCAP Specific IgE Negative control as quotient, type 100
for Range maximum.
NOTE!
If expected mean value, expected standard deviation and maximal
allowed standard deviation have been defined, the Range maximum will be
overridden.
Log start date
The following step describes how you can change the log start date.
1.
In Edit QC mode press
11
.
2.
Press
yes
.
3.
Delete or type over the
old log start date. Type the new
log start date and press
enter
to confirm this step.
Lot number
The following step describes how you can change the lot number (used mainly
for Phadia controls).
1.
In Edit QC mode press
12
.
2.
Press
yes
.
3.
Delete or type over the
old lot number. Type the new
lot number and press
enter
to
confirm this step.
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
1.Status?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
11.Log start date?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
Log start date: 060410
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
1.Status?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
12.Lot number?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
Lot number: 14371956
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
Range maximum: 4.2
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
10.Range maximum?
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...