Phadia 100 - User Manual
Routine Assay Run
12-3501-20/12
Page 2.43
2
3
4
5
6
2
13
Warnings that may prevent entering
RUN IS NOT PREPARED
There is no currently prepared run.
PROCESS IS RUNNING
There is an assay run being processed. You may prepare a new run, but you
cannot load and start.
4.
Press
yes
to load the
samples.
Remove the caps on the reagents/samples before placing them in the Sample
Carousel.
Make sure that there is enough amount to be withdrawn.
Check for air bubbles or film on the surface of the reagents/samples. If there are
air bubbles or film, pinch them away. Otherwise the pipette sensor will not
detect the surface properly.
5.
Check that the Washing,
Rinse and Waste bottles are
loaded. Press
yes
to confirm
this step.
Make sure the volume is enough and that the tube connection is pressed down
until you hear the clicking sound.
Empty the ImmunoCAP Waste Container.
1.3.Load and Start
1.Load Samples?
1.3.1.Load Samples
WASH/RINSE/WASTE OK?
NOTE! When running ImmunoCAP Specific IgA,
ImmunoCAP Specific IgG or ImmunoCAP Specific
IgG4 with a full sample carousel, 1 l of Washing
Solution is not enough. Fill up the bottle with
Washing Solution.
WARNING! Waste bottle, ImmunoCAP waste
container and process chamber may be
contaminated by potentially infectious,
biohazard material - take appropriate actions.
Wear protective gloves!
WARNING! Washing Solution may cause
sensitization by skin contact. Wear protective
gloves!
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...