Phadia 100 - User Manual
Maintenance
12-3501-20/12
Page 11.17
2
5
6
11
6.9.3.1.FluoroC Run Target:DW_
LOAD 1890 ul FLUOROC IN POS F
6.Install/Service
1.Login/Logout?
6.Install/Service
9.Maintenance?
6.9.Maintenance
3.FluoroC?
6.9.3.FluoroC
1.FluoroC Run?
FluoroC Run
The FluoroC run is used for service and trouble shooting, and checks the
performance of the fluorometer.
To run FluoroC you need FluoroC Solution and Stop Solution. The FluoroC vial
must be at room temperature before the run. Use only new FluoroC vials in each
run.
Make sure that the Washing Solution and Rinse Solution Bottles are filled and
connected to the instrument.
The FluoroC run takes approximately seven minutes regardless of the warm-up
time.
1.
In the Prepare Run mode
select the number
6
key.
2.
Press the
yes
key.
3.
In the Install/Service
mode select the number
9
key.
4.
Press the
yes
key.
5.
In the Maintenance mode
select the number
3
key.
6.
Press the
yes
key.
7.
Press the
yes
key.
8.
Check the target value of
the FluoroC and load the
FluoroC vial in the given posi-
tion.
6.9.Maintenance
1.Daily Maintenance?
Phadia 100 Ver.3.0
07:55:06
6.Install/Service?
Phadia 100 Ver.3.0
07:55:06
1.Prepare Run?
WARNING! Washing Solution may cause
sensitization by skin contact. Wear protective
gloves!
FluoroC target
value (on the vial)
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...