Printouts
Phadia 100 - User Manual
P a g e 1 0 . 6
12-3501-20/12
Laboratory Report
The laboratory report contains the results from one assay run. It is automati-
cally printed after an assay run has been processed. It can also be printed for any
stored assay run by selecting
3.2.2.Specific Run
or
3.2.3.Specific Run, select
Calibrators
.
You can customize the laboratory report in
6.4.System Parameters
as follows:
• Include calibration curve You can decide if you want the stored calibration
curve graph for a run to be printed or not. How-
ever, if calibrators were processed in the run the
calibration curve data will always be printed.
• Include calibrator data
You can decide if you want the stored calibrator
data for a run to be printed or not. However, if
calibrators were processed in the run, the calibra-
tor data will always be printed.
• Unknown data order
Unknown data can be listed by sample or by
method/test.
• Number of copies
You can specify the number of laboratory report
copies to be automatically printed after each run.
You can select any number between 0 and 3.
Cont...
LABORATORY REPORT
Run number:
18
Run created:
090508 16:15
Process started:
090509 08:23
Process ended:
090509 11:19
Run origin:
Instrument
ImmunoCAP
Specific IgE 0-100
Calibration code: AA
Monthly cal:
NO
CALIBRATION INFORMATION
Calibration number: 3
Run date, time:
090508 08:23
Calibration code: AA
Acceptance:
OK
Calculation type: Rodbard 5-parameters
Run sequence number
Date and time of the first entered
request
Date and time when the assay
run process was started
Date and time when the assay
run process was ended
Run origin:
Instrument The run was created
in the instrument
PC Remote
The run was created
on a PC using IDM
Each section begins with the
method name and contains all
the results of that method
Calibration code for this method
and run
Monthly calibration run:
NO
The calibration curve is
included in the assay run
YES
The calibration curve is not
included in the assay run
Date and time when the Calibra-
tion curve was run
Acceptance of the calibration
curve used:
OK
Calibration curve
accepted
FLAGGED
Calibration curve
flagged
NOT OK
Calibration curve
not accepted
The sequence number used for
the calibration curve within the
method
An error message is printed here
if a Calibration Curve Error has
occured
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...