Messages that can occur
NO ACTIVE CALIBRATORS, ADD TO RUN?
There is no stored calibration curve for the method.
NO TEST POS LEFT, Skip Sample ID?
There is no position left for ImmunoCAP/EliA Well. If you press
yes
the sample
identity will be removed.
NO TEST POS LEFT, Skip Test?
There is no position left for ImmunoCAP/EliA Well. If you press
yes
the test will
be removed.
NO TEST POS LEFT, Skip Panel?
There are not enough ImmunoCAP/EliA Well positions left. If you press
yes
the
test panel will be removed.
NO SAMPLE POS LEFT, Skip Sample ID?
There is no position left for sample. If you press
yes
the sample identity will be
removed.
NO SAMPLE POS LEFT, Skip Test?
There is no position left for sample. If you press
yes
the test will be removed.
NO SAMPLE POS LEFT, Skip Panel?
There is no position left for sample. If you press
yes
the test panel will be
removed.
PREDILUTED 1:10?
Press
yes
if the sample is prediluted ten times (this example).
INSTRUMENT DILUTION 1:10?
Press
yes
if the test is to be diluted ten times (this example). The dilution is
carried out by the instrument.
TEST ALREADY ENTERED
The test name has already been entered for the sample.
TOO MANY METHODS AT THE SAME TIME
The number of methods possible to enter was exceeded. Maximal number of
methods that can be entered in one run is four (4). If ImmunoCAP and EliA are
entered the message METHOD NOT COMPATIBLE should show up.
ENTERED TEST DOES NOT EXIST
The entered test name is not defined in the currently available methods.
WRONG BAR-CODE IDENTITY
Bar-code was read, but could not be recognized as a test type. Check that you
have not read the EAN code instead of the article bar-code.
METHOD NOT COMPATIBLE
El-G method is not compatible with methods using ImmunoCAP.
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...