Assay Run Quality Control
Phadia 100 - User Manual
P a g e 6 . 8
12-3501-20/12
Print Calibrators
4.Quality Control
1.Print Log?
4.Quality Control
3.Print Calibrators?
4.3.Print Calibrators
Method: ImmunoCAP Specific IgE 0-100?
4.3.Print Calibrators
[sIgE]
1.Specific Calibrators?
4.3.1.Specific Calibrators
[sIgE]
sIgE calibrators: 090405 11:13 153?
You can print calibration curve and data for selected calibrators.
1.
In Prepare Run mode
press
4
.
2.
Press
yes
.
3.
In Print Log press
3
.
4.
Press
yes
.
5.
Press
no
until the method
you want appears, then press
yes
.
Messages that can occur
CANNOT SELECT THIS METHOD
The selected method cannot be used for this operation.
6.
Press
yes
.
Warnings that may prevent entering
NO CALIBRATORS EXIST
There are no stored calibrators for the selected method.
7.
Select the calibrators you
want by pressing
no
until the
ones you want turn up, then
press
yes
. The selected calibra-
tors curve and data are printed.
Phadia 100 Ver.3.0
14:10:17
1.Prepare Run?
Phadia 100 Ver.3.0
14:10:17
4.Quality Control?
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...