Methods, Test Panels, Article Management
Phadia 100 - User Manual
P a g e 9 . 2 2
12-3501-20/12
5.2.9.2.Edit QC
In the Edit mode it is possible to change status, identity, test name, replicate,
response, expected conc. and s, limit s, range minimum and maximum as well
as log start date and number.
Warnings that may prevent entering
NO QUALITY CONTROLS EXIST
There is no QC defined for the method.
Select the Quality Control you want to edit.
Select the parameter you want to edit.
5.2.9.2.1.Status
Select ACTIVE, MANUAL or INACTIVE: Confirm with
yes
.
Active means that it will be added in all assay runs. Inactive means that it will
not be added. Manual means that it will be added if you decide so when an assay
run is prepared.
5.2.9.2.2.Identity
Type identity. Confirm with
enter
.
Messages that can occur
SAME QUALITY CTRL IDENTITY ALREADY EXISTS
The identity is already used by another QC.
It is possible to use the same vial (= same identity) for different tests.
5.2.9.2.3.Test name
Type Test name. Confirm with
enter
.