Error Handling
Phadia 100 - User Manual
P a g e 1 2 . 4 8
12-3501-20/12
Example from a Laboratory Report:
Make sure that all reagents and samples are in their right positions according
to the Distribution List and contain sufficient fluid according to the Consumption
List.
If the error remains contact your local Phadia representative.
Quality Control HIGH
What can cause the problem:
Evaporation from QC bottle causing to high concentration in bottle. Problem
with Stop Solution pipetting. Wrong QC limits or wrong QC bottle used. Wrong
ImmunoCAP/EliA Well dispensed. For some
EliA
QCs (specially dsDNA) and
Gliadin
QC results is sensitive to concentration of Washing Solution. For
EliA
it is possible that wrong Lot Specific Code has been used.
EliA
QCs are designed
for single use only.
Corrective actions:
No corrective actions proposed in Action List , except for
EliA
methods.
Follow instructions in Action List in order to identify or eliminate the errors. For
additional information see Explanation of Actions, number
17, 25, 26.
If the error remains contact your local Phadia representative.
Quality Control LOW
What can cause the problem:
Pipetting problem caused by that Washing and Rinse Solution distribution is
affected. Too small volume QC loaded (big dead volume in QC bottle) or air
bubbles in QC bottle.
EliA
QCs are designed for single use only. Wrong QC limits
or QC bottle used. Wrong ImmunoCAP/EliA Well dispensed. For
EliA
it is
possible that incorrect Lot Specific Code has been used or wrong storage
conditions of EliA Wells. For some QC results (specially
EliA dsDNA
) concen-
tration of Washing Solution is critical.
QUALITY CONTROLS
NAME----RESP--%CV--CONC--%CV--JUDGED
QC1
206
0.46
L O W
(t3)
QC1
203
0.45
LOW
(d1)
NAME----EXP CONC--S--LIMIT--MIN--MAX
QC1
5.00
0.00
3.0 2.70 6.30
QC1
0.00
0.00
1.0 2.00 5.00
OK, HIGH or LOW is printed here
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...