Phadia 100 - User Manual
Routine Assay Run
12-3501-20/12
Page 2.49
2
3
4
5
6
2
13
DAILY MAINTENANCE NOT PERFORMED
Perform Daily Maintenance.
WEEKLY MAINTENANCE NOT PERFORMED
Perform Weekly Maintenance.
MONTHLY MAINTENANCE NOT PERFORMED
Perform Monthly Maintenance.
TEMPERATURE SENSORS DIFFER
For corrective actions see Action list.
18.
To start the process press
enter
.
As soon as the tempera-
ture has reached 37ºC ± 0.5ºC
the instrument will start the
process.
A message tells you that the
process has started.
19.
If there were calibrators added to the run, you will get a message about the
status of the calibration. If status is OK, the calibration curve is stored and
available for use in routine runs. If status is NOT OK, the calibration curve is
stored but not available for use in routine runs.
20.
When the run is processed,
there will be a message telling
you that the process is com-
pleted and a report is printed.
21.
Empty Waste bottle and ImmunoCAP waste container after each run.
22.
Daily maintenance has to be performed at the end of the day. See
chapter 11, Maintenance, for further instructions.
1.3.4.Start Process
PRESS ENTER TO START PROCESS
1.3.4.Start Process
PROCESS STARTED
PROCESS COMPLETED
EMPTY AND CLEAN INSTRUMENT
WARNING! Waste bottle, ImmunoCAP waste
container and process chamber may be con-
taminated by potentially infectious, biohazard
material - take appropriate actions. Wear pro-
tective gloves!
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...