Assay Run Quality Control
Phadia 100 - User Manual
P a g e 6 . 1 2
12-3501-20/12
4.5.1.Export Quality control-logs
INSERT MEDIA AND PRESS ENTER
4.5.ImmunoCAP<->PC
1.Export Quality control-logs?
4.5.1.Export Quality control-logs
Method: ImmunoCAP Specific IgE 0-100?
4.4.3.Edit Calibrator Point
[sIgE]
Point: CAL-0.35?
4.4.3.Edit Calibrator Point
[sIgE]
Response: (2) 58.3_
4.
Press
no
until the calibra-
tor point you want appears and
then press
yes
.
5.
Type the new response of
the calibrator point and press
enter
.
If you want to recall the original response press
enter
and
shift
at the same time.
If you want to exit without any changes done press
mode
. The number of
replicates are displayed.
Export Quality Control-Logs
1.
In the Export Quality con-
trol-log mode press
yes
.
2.
Press
no
until the method
you want appears, then press
yes
.
Warnings that may prevent entering
NO QUALITY CONTROL-LOGS EXIST
There are no QC logs for the selected method.
CANNOT SELECT THIS METHOD
The selected method cannot be used for this operation.
3.
Insert a USB memory or a
diskette, or connect an external
USB hard drive, and press
en-
ter
.
NOTE! Use only one of the USB ports on the
front at a time.
The quality control-log for the selected method will be exported.
Summary of Contents for Phadia 100
Page 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Page 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Page 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Page 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Page 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Page 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Page 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...